Alaris™ GP (Guardrails™) Volumetric Pump
Pump Configuration available via the Alaris Editor Software
BDDF00535 Issue 1
30/56
Large Volume Pump Configuration Settings
AIL Limit The single bubble Air-in-line alarm setting.
Bolus Mode Controls whether or not the Pump allows bolus delivery method.
Bolus Rate Default The default value for bolus rates.
Bolus Rate Max The maximum permissible bolus rate.
Bolus Volume Max The maximum permissible bolus volume.
Occlusion Alarm Pressure The default occlusion limit.
Pressure Max The maximum permitted occlusion limit.
Infusion Rate Max The maximum permissible infusion rate.
KVO Rate The maximum infusion rate when performing KVO delivery.
Near End of Infusion Point The point defined in terms of remaining time at which the infusion is deemed to be Near end i.e.
nearly complete.
Primary VTBI Max The maximum VTBI for primary infusions.
Prime Rate The rate at which priming is performed.
Prime Volume Max Determines the maximum volume that can be purged as part of a given prime operation.
Secondary Infusion Allows the use of a secondary infusion (Piggyback) in the same channel.
Sec. Infusion Rate Max The maximum permissible infusion rate for secondary infusions.
Sec. VTBI Max The maximum permissible setting for the Volume To Be Infused for secondary infusions.