Alaris™ GP (Guardrails™) Volumetric Pump
Specifications
BDDF00535 Issue 1
48/56
Specifications
Electrical Protection
Class I, Type CF (Defibrillation-proof)
Electrical/Mechanical Safety
Complies with IEC/EN60601-1 and IEC/EN60601-2-24.
Electro Magnetic Compatibility (EMC)
Complies with IEC/EN60601-1-2 and IEC/EN60601-2-24.
Electrical Safety
Typical earth leakage current 78µA.
Typical Enclosure Leakage Current (Normal Condition) = 0µA
Typical Protective Earth Resistance = 32mOhms
The above measurements are for guidance only, IEC/EN60601-1 limits are defined below:
Earth Leakage Current (Normal Condition) <= 500µA
Enclosure Leakage Current (Normal Condition) <= 100µA
Protective Earth Resistance <= 200mOhms
Classification
Continuous mode of operation, Portable Equipment
AC Power Supply
100 - 230 VAC, 50 - 60Hz, 60VA (Maximum).
Fuse Type
2 X T 1.25 A, slow blowing.
Dimensions
148mm (w) x 225mm (h) x 148mm (d). Weight: approx. 2.5kg (excluding power cable).
Protection against fluid ingress
IP33 - Protected against direct sprays up to 60° from the vertical.
Environmental Specifications
Condition Operating Transport & Storage
Temperature +0°C - +40°C -20°C - +50°C
Humidity 20% - 90%* 15% - 95%*
Atmospheric Pressure 700hPa - 1060 hPa 500hPa - 1060hPa
*Non condensing.
Battery Specifications
Rechargeable NiMH (Nickel Metal Hydride). Automatically charges when the Pump is connected to AC power.
Battery Charging - 2.5 hours to 95%.
Battery Life
Infusion Rate Mean Time To Battery Empty from fully charged
25ml/h 5.8 hours
125ml/h 5.65 hours
1200ml/h 3.4 hours

Table of Contents