If the Network Communications Board is installed, all of the Area Alarm details
can be browsed by using the Web page of the Master Alarm or the Area Alarm.
With a Network Communications Board installed, the Area Alarm memory can
be downloaded.
The Area Alarm Modules have gas-specific, color-coded labels for both USA and
ISO (see table 1-9 on page 1–35).
Table 1-9. Area Display Module Gas Labels
a.Text is both black and white in a striped fashion.
An additional label space is provided at the base of each module, for field use, to
note the location that the module is monitoring.
Gas name Text Color Background Color Use
Oxygen White Green USA
Air Black Yellow USA
Vacuum Black White USA
Nitrogen White Black USA/ISO
Nitrous Oxide White Blue USA/ISO
Evacuation
(WAGD)
White Purple USA
Carbon Dioxide Black Gray USA/ISO
Helium White Brown USA/ISO
CO
2
/O
2
Gray/green White USA
Air
a
Black and White White and Black ISO
Oxygen Green White ISO
Vacuum Black Yellow ISO
Air
a
Black and White White and Black French
Oxygene Green White French
Protoxyde D’Azote
(N
2
O)
White Blue French
Succion (vacuum) Black Yellow French
Azote (N
2
)
White Black French