EasyManua.ls Logo

BeaconMedaes MedPlus TotalAlert - Page 56

BeaconMedaes MedPlus TotalAlert
376 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table 1-16. MedPlus TotalAlert® Alarm Network Conversion Kits
Existing Alarm Type
Existing Alarm
Rough-In Box Sizes
MedPlus TotalAlert®
Alarm Network
Conversion Kit
MedPlus® Medical Gas Alarm
(Digital) from BeaconMedaes, Hill-
Rom or Medaes
Master Alarm (30-point maximum)
14" x 9" (355.6 mm x
228.6 mm)
Master Alarm 6819-A9
Area Alarm (3-gas maximum)
14" x 9" (355.6 mm x
228.6 mm)
Three-Module Area
Alarm 6819-A10
Area Alarm (6-gas maximum)
24" x 9" (609.6 mm x
228.6 mm)
Six-Module Area
Alarm 6819-A11
Master Alarm plus Area Alarm (30-point
maximum master and a 3-gas maximum
area)
14" x 9" (355.6 mm x
228.6 mm)
Combination Alarm
6819-A13
Area Line Pressure Alarm (0321) from
BeaconMedaes, Hill-Rom, Medaes, or
Ohmeda (analog alarm having a beige
front panel)
Area Line Pressure Alarm (6-gas maxi-
mum)
12 7/8" x 8 3/8"
(327.03 mm x 212.73
mm)
Six-Module Area
Alarm 6819-A11
Integrated Alarm System or Integrated
Alarm System II from Medaes, or
Ohmeda
Master Alarm (30-point maximum)
7 ½" x 9 ½" (190.5
mm x 241.3 mm)
Master Alarm 6819-A9
Integrated Alarm System I or Integrated
Alarm System II (3-gas maximum)
7½" x 9½" (190.5 mm
x 241.3 mm)
Three-Module Area
Alarm 6819A10
Integrated Alarm System I or Integrated
Alarm System II (6-gas maximum)
7½" x 9½" (190.5 mm
x 241.3 mm)
Six-Module Area
Alarm 6819-A11
Integrated Alarm System I or Integrated
Alarm System II (30-point maximum
master and a 3-gas maximum area)
7½" x 9½" (190.5 mm
x 241.3 mm)
Combination Alarm
6819-A13
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728) Page 1-43
Specifications
Chapter 1: Introduction
1

Table of Contents