4
Insert power unit into AC outlet, which shall be easily accessible; after few
seconds the balance will automatically switch on.
Balance Warm-up:
Wait 8 hours from switch one for Analytical balanaces (0,0001g)
Wait 30 minutes for other models
Then calibrate the balance using the appropriate mass (if supplied), following
instructions (par.7).
It is strongly suggested not to disconnect the balance from the electrical
mains and to use the ON/OFF button to stand-by the balance when you
finished using it
Calibrate again the balance every time it’s moved to another place
Check balance calibration periodically.
We recommend not dropping heavy objects on balance pan, in order to avoid
damages.
Assistance service must be effected by specialized staff and the spare parts used
must be original. Therefore, it is necessary to apply to the reseller who sold the
equipment.
2 Storage
.
Storage temperature: +5 °C…+40°C
Storage humidity 45% - 75%.
Keep package in the case the balance should be sent back to the factory for service.
Disconnect the cables and the accessories for avoiding damages during
transportation.
Do not place the balance in extreme temperature and humidity condition and avoid
the balance to take strong hit.