28 Error codes
ERROR MESSAGE
ON DISPLAY
MEANING POSSIBLE SOLUTIONS
ERR01
Weight not stable after operation
of tare
Protect the balance from air
flows or from vibrations of
the working table
ERR02
impossible to start the calibration
due to instability of the balance
Protect the balance from air
flows or from vibrations of
the working table.
ERR03
calibration weight not correct or
balance unstable
Calibrate with correct
weight or protect the
balance from environment
disturbs
ERR04
weight of samples for the
piececounting function not
adequate or unstable
Select a bigger number of
samples or protect the
balance from vibrations.
ERR05
impossible to print / weight
unstable
Protect the balance from
environment disturbs
ERR06
Weight cannot get stable in
density mode
Protect the balance from
environment disturbs
ERR07 error in inserting the data
ERR08 Anomaly on autocalibration motor Contact service staff
“UNLOAD”
weight loaded on the pan or pan
not positioned properly
Remove the weight from
the pan or position properly
the pan and underpan.
“CAL But”:
the balance requires to be re-
calibrated
Unload weights, if any, on
the pan, and press the CAL
button
Over-range condition
Unload the weights loaded
on the pan
Under-range condition
Place properly pan and
underpan