EasyManuals Logo
Home>Bender>Measuring Instruments>ISOMETER iso1685D Series

Bender ISOMETER iso1685D Series User Manual

Bender ISOMETER iso1685D Series
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #38 background imageLoading...
Page #38 background image
Device communication via the BMS busDevice communication via the BMS bus
iso1685DP_D00272_03_M_XXEN/08.2017
38
Wiring
The following type of wiring is recommended for the RS-485 network:
Shielded cable, core diameter ³ 0.8 mm
(e.g. J-Y(St)Y 2x0.8), shield connected to earth (PE) on one end.
Connection to terminals A and B.
The number of bus nodes is restricted to 32 devices. If more devices are to be connected,
Bender recommends the use of a DI1 repeater.
10.3 BMS protocol
This protocol is an essential part of the Bender measuring device interface (BMS bus pro-
tocol). Data transmission generally makes use of ASCII characters.
Interface data are:
Baud rate: 9600 baud
Transmission: 1 start bit, 7 data bits, 1 parity bit, 1 stop bit (1, 7, E, 1)
Parity: even
Checksum: sum of all transmitted bytes = 0 (without CR and LF)
The BMS bus protocol works according to the master-slave principle. Only one master
may exist in each network. All bus devices are identified by a unique BMS address. The
master cyclically scans all other slaves on the bus, listens to their signals and then carries
out the corresponding commands.
A device receives the MASTER function by assigning Bus address 1 to it.
10.4 BMS master
A master can query all measured values, alarm and operating messages from a slave. If
bus address 1 is assigned to a device, this device automatically represents the master, i.e.
all addresses between 1 and 150 are cyclically scanned via the BMS bus for alarm and op-
erating messages. If the master detects incorrect answers from a slave, the fault message
"Fault RS-485" will be output via the BMS bus.
DI1
1
32
33
64
Fault causes may be:
Addresses are assigned twice
A second master exists on the BMS bus
Interference signals occur on the bus lines
A defective device is connected to the bus
Terminating resistors are not activated or connected
10.5 Commissioning of an RS-485 network with BMS protocol
Interconnect terminals A and B of all bus devices in one line
Switch the terminating resistors on at the start and the end of the RS-485 network. If
a device at the end of the bus is not terminated, connect a 120  resistor to termi-
nals A and B
Switch the supply voltage on
Assign the master function and address 1 to a bus-capable device
Assign addresses (2…90) to all other bus devices in consecutive order
10.6 Setting BMS address
Set the BMS address ((1)2…90) in the device menu via the path:
Device settings/Interface/BMS/BMS Address.
The ISOMETER® cannot switch on a potential termination at the BMS bus.
Even though this is not expected to cause communication problems, the
ISOMETER® should be operated as BMS slave if possible (BMS address > 1).
If no other device with master capabilities is available on the bus, the
ISOMETER® can be set to master (BMS address 1).
Before the ISOMETER® takes over the backup master function after being
switched on, it waits to see if another master connects itself to the system.
Waiting period: BMS address minus 1 = waiting period in minutes.
Example: The iso1685DP has BMS address 3. It waits 3 minus 1 minutes
(= 2 minutes) for a master to connect.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bender ISOMETER iso1685D Series and is the answer not in the manual?

Bender ISOMETER iso1685D Series Specifications

General IconGeneral
BrandBender
ModelISOMETER iso1685D Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals