EasyManua.ls Logo

Berthold Uni-Probe LB 491 - Reading-In Pulse Rates

Default Icon
622 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
38478BA2B
6 – 554 27.5.08
6 Explanations Volume 6
6.3 Reading-in Pulse Rates
For each calibration you have to read-in or enter count rates. In the
Uni-Probe LB 491 User’s Manual, Volume 6 there are several refer-
ences to this procedure.
Read or enter count rate?
The count rate can be read-in or entered. Typically, the count rate
is read in. If the count rate has already been determined in the
past, or in another manner, then you can enter the count rate man-
ually.
6.3.1 Proceed as follows
Select READ and confirm the selection with OK.
READ means the current count rate is read in.
ENTER allows you to enter a count rate manually.
Reading the count rate:
If necessary, change the value for the averaging time.
The standard value 0 selects an averaging time of 30s, which suf-
fices in most cases. This means that the process is over after
30 seconds.
Enter a value >0 to determine another averaging time in seconds.
The longer this time, the more accurate the averaging.
IMPORTANT
Do not enter a number that is too high by mistake!
The process can be stopped only by turning off the Uni-Probe.
Click OK to start reading in the count rate.
IMPORTANT
During averaging the density must not change.
The count rate reading-in process has started.
If the process has been started with 0, it will take 30s until the
count rate is read in. Otherwise, the duration of this process corre-
sponds to the averaging time entered.
“Reading” means that all incoming count rates are added up. How-
ever, a count rate value which is standardized with respect to one
second is always displayed. Thus, an averaged count rate is calcu-
lated.

Table of Contents

Related product manuals