Volume 1 5 Repair, Maintenance and Service
1
5.10 Replacing Fuses
Position Two fuses (16) are located on the power supply board (14).
Fig. 5-6 Position of power supply board and fuses in the device
Technical data of fuse Type of fuse: Berthold part #24586.
IMPORTANT
For power supply 95V to 250V
AC
and 24V
AC/DC
, use only 1.25A/
250VAC/slow-blow fuses with UL license!
Proceed as follows to replace fuses Dismantle the electronics module as described on page 1–91.
Exchange fuses.
Install the electronics module in die Uni-Probe as described on
page 1–92.
This completes the replacement of the fuses.