Volume 1 5 Repair, Maintenance and Service
1
5.6.1 Dismantling the Electronics Module
Disconnect the Uni-Probe, possible connected relay contacts
and all in- and outputs from power.
Unscrew housing cover.
Remove grounding cable (1).
Unscrew two nuts M8 (2).
Fig. 5-2 Dismantling the detector electronics
Carefully pull out electronics module horizontally together with
the crystal-multiplier assembly.