6 Libra S21/S22, English Issue 02 - 04/2004
Basic Modes (1)
Absorbance (1.1)
Absorbance mode measures the amount of light that has passed through a sample
relative to a blank (this can be air). The procedure is as follows:
Enter appropriate wavelength and press OK (F3)
Insert reference and press green run key. The cell changer, if fitted,
automatically moves to position 2 and displays the result for the reference
measurement (0.000)
Xenon lamp based instruments are “press to read”, whereas deuterium /
tungsten lamp instruments measure continuously. Thus to monitor sample
stabilisation, the simple kinetics mode must be used
This reference value is used for subsequent samples until changed
Insert samples as required and press (repeat as necessary)
To go back and change the wavelength press Method (F1)
% Transmission (1.2)
Transmission mode measures the amount of light that has passed through a sample
relative to a blank (this can be air), but displays the result as a percentage. The
procedure is as follows:
Enter appropriate wavelength and press OK (F3)
Insert reference and press green run key
This reference value is used for subsequent samples until changed
Insert samples as required and press (repeat as necessary)
To go back and change the wavelength press Method (F1)
Factor Concentration (1.3)
Concentration mode is used when a conversion factor is known, and converts the
absorbance measurement for a sample at a specific wavelength into a concentration,
by a simple multiplication of absorbance x factor. The procedure is as follows:
Enter appropriate wavelength and press OK (F3)
Enter known factor (range 0.01-9999) and press OK (F3)
To enter a negative factor press – (F1); the reference should have a higher
absorbance than the samples
Insert reference and press green run key
This reference value is used for subsequent samples until changed
Insert samples as required and press (repeat as necessary)
To go back and change the wavelength or factor press Method (F1)