2 Libra S21/S22, English Issue 02 - 04/2004
Unpacking, Positioning and Installation
• Inspect the instrument for any signs of damage caused in transit. If any damage
is discovered, inform your supplier immediately.
• Ensure your proposed installation site conforms to the environmental conditions
for safe operation:
Indoor use only
Temperature 10°C to 40°C
Maximum relative humidity of 80 % up to 31°C decreasing linearly to 50 %
at 40°C
• The instrument must be placed on a hard flat surface, for example a laboratory
bench or table, which can take its weight (13 kg) such that air is allowed to
circulate freely around the instrument.
• Ensure that the cooling fan inlets and outlets are not obstructed; position at least
2 inches from the wall.
• This equipment must be connected to the power supply with the power cord
supplied and must be earthed (grounded)
. It can be used on 90 - 240V supplies.
• Switch on the instrument and check that the display works (see Operation).
• To enter laboratory name, operator name, instrument asset number details, and
current date/time, refer to System Utilities.
If this equipment is used in a manner not specified or in environmental conditions
not appropriate for safe operation, the protection provided by the equipment may be
impaired and instrument warranty withdrawn.