Issue 02 - 04/2004 Libra S21/S22, English 9
Simple Kinetics (2.2)
Simple kinetics studies to investigate the shape of an assay curve can be readily
performed. The wavelength of interest is entered together with the time interval at
which absorbances are to be read: the results are displayed graphically, simulating a
chart recorder output. The procedure is as follows:
Enter appropriate wavelength and press OK (F3)
Select time units; seconds (1) or minutes (2)
Enter the duration of the assay and press OK (F3)
Enter the time interval; minimum 2, maximum 60 seconds
Select if the actual absorbance time data should be printed with the results
Insert reference and press green run key
This reference value is used for subsequent samples until changed
Insert samples as required and press (repeat as necessary)
To see the assay on the whole display, press Data (F3); to return press OK (F3)
Data points can be viewed by moving the cursor (F2 and F1); this enables
the identification of slope start and end times, for example
To go back and change the parameters press Method (F1)
NOTE
This mode should be used to check sample stabilisation prior to kinetics studies, for
example, since the xenon lamp is not a continuous output source (unlike deuterium
and tungsten lamps).