Issue 02 - 04/2004 Libra S21/S22, English 17
Download to Spreadsheet
Results can be downloaded directly to Excel when the PC has the Spreadsheet
Interface Software installed (80-2112-23) and the two are linked with the serial cable
(80-2105-97); detailed instructions are supplied with the software. Thus absorbance
/ wavelength data comprising a scan, for example, can be picked up as columns of
numbers and converted to a more conventional graph using the spreadsheet; results
can then be formatted or manipulated as appropriate prior to inclusion in reports or
archiving / saving to hard disk.
Results from all modes of use on the instrument can be output in this way. Output is
automatic when the
key is pressed.
Messages
Most messages are self-explanatory and relate to use of the instrument.
Others relate to the calibration of the instrument on switch on:
This instrument has
failed 1 or more GLP
tests
One or more of the parameters tested for during GLP calibration
is out of specification (see Appendix). You can accept this
status and continue to use the instrument as normal, but you
may to contact your local service engineer
Failed to find Abs Failed to calibrate properly; contact local service engineer
Failed to find Ref 1 Failed to calibrate properly; contact local service engineer
Failed to align filters Failed to calibrate properly; contact local service engineer
Failed to align grating Failed to calibrate properly; contact local service engineer