Issue 02 - 04/2004 Libra S21/S22, English 15
GLP (4)
Refer to Appendix for more information. This option determines whether GLP is on
or off in terms of printing and reporting the results; the calibration interval for GLP,
however, is always on and can be done automatically at pre-defined time intervals
(always on, daily, weekly, monthly, quarterly). If GLP is on, the results are printed
automatically after calibration; they can also be printed on demand using Print (F2)
on the System page. Note that the GLP print out will show the date for when the full
calibration was done (“Calibrated”), and that this can be different to the date of
instrument operation (“Date”); this is shown on the example below. If the date is the
same, “Calibrated” shows the time that it was done instead.
Press “More…” (F3) on the system page to view the GLP results on the instrument
display.
Libra S22 GLP Report
Instrument Libra S22
Operator A T Dadd
Date 22 March 2002
Time 10:00:17
Serial No. 79500
Version 6090 V1.0
Calibrated 22 March 2002
Instrument Life 25.6 Hours
Service 22 March 2002
Bandwidth
(2.0 – 3.0nm) 2.9 PASS
Wavelength Accuracy
881.9nm (± 1 nm) 881.9 PASS
Absorbance Accuracy
220nm (1.763 – 1.781A) 1.772 PASS
340nm (1.633 – 1.665A) 1.649 PASS
500nm (1.477 – 1.491A) 1.484 PASS
Stray Light
220 nm (<0.05%) 0.021 PASS
Language (5)
Select language for the display and print out.
Service (6)
This is for accredited service engineers only and requires the entry of a pass code.