Rev A N-21
US
ER
I
D
U
p to 600 User ID numbers with 1 – 10 digits may be programmed into the meter.
T
he User ID may alternately be entered using the external Bar Code Scanner (optional).
An additional security feature of the User ID is the partial masking of the ID Number on
displays and printouts. This prevents unauthorized individuals from detecting and using an
existing User ID number. When the Supervisor Code Chip module is installed in the meter,
this feature is overridden and all User ID digits are displayed.
NE
W
US
ER
ID
1. From the list of programmable parameters, choose US
ER
ID
using keys.
2. Press the key.
3. Select
NE
W
US
ER
ID using keys.
4. Press the key.
5. Type in a number from 1 – 9999999998 (10 digits). To correct, press the key
to clear the entire number, or select the incorrect digit by using the
keys. Then type in the correct number.
6. Press the key.
7. On the next screen, use the keys to select an expiration date to the
right of
USER ID STATUS: 6 MONTHS, 12 MONTHS, VALID or EXPIRED.
8. Press the key to save changes.
The Triage module MeterPr
o r
ecognizes leading 0’s as a unique character. Therefore,
‘4341’ will be recognized separately fr
om ‘04341.’
Using the barcode scanner or alpha mode will allow the Triage module MeterPro to accept
alpha characters and the following symbols: # ( ) * - . / \
MAIN MENU
SET PARAMETERS
USER ID
Note: USER ID
9999999999
(10-9’s) is a default
setting displayed when
the
USER ID BYPASS
is on or when a test
is run with the
Supervisor Code Chip
module installed.
Note: If you
type in a
User ID number that
has already been
assigned, the
message on the
screen states:
USER ###
ALREADY EXISTS
.
Press the
ENTER key
to go back and type
in a different number.