34
! Using COMP (Compare) Function
“COMP”
function compares the measurement inputs with the pre-set upper and
lower limits. The compare function calculation expression is based on counts
without decimal point.
HI: Measurement value > High (HI) limit value
LO: Measurement value < Low (LO) limit value
PASS: High limit value
Measurement value
Low limit value
When
“COMP”
function is enabled, the actual measurement value will be shown in
primary display and a comparison result
“HI”, “LO”
, or “
PASS”
will be shown in
secondary display.
! Operation Procedures
then
,
to enable the
“COMP”
function. Press
then
again to disable the compare function.
! Setting a Compare Limit Value
Using the following procedure to set the high and the low limit values for
“COMP
function:
then to enter the high limit set-up mode.