78
CONFigure
[:SCALar]
:VOLTage
:DC [<numeric value>] [ [ ,@1] | ,@2 ]
:AC [<numeric value>] [ [ ,@1] | ,@2 ]
:ACDC [<numeric value>] [ ,@1]
:DCAC [<numeric value>] [ ,@1]
Description : Set the primary and secondary display to voltage measurement function for
the DCV, ACV or AC+DC mode. The <numeric value> parameter is used to
configure the measurement range. [ [@1] | ,@2 ] has the same syntax as
SCPI <channel_list> syntax. Use @1 to select the primary display function,
and @2 to select the secondary display function.
Parameter : <numeric value> ; For volts range : 0.12, 0.4, 1.2, 4, 12, 40, 120, 400,
750, 1000 V.
[ [ ,@1] | ,@2 ] ; For the primary display and secondary display setting.
Example : CONF:VOLT:DC ; Set the primary display to DC volts .
CONF:VOLT:AC ,@2 ; Set the secondary display to AC volts .
CONF:VOLT:ACDC 0.12,@1 ; Set the primary display to AC+DC
volts 120mV range .
CONFigure
[:SCALar]
:CURRent
:DC [<numeric value>] [ [ ,@1] | ,@2 ]
:AC [<numeric value>] [ [ ,@1] | ,@2 ]
:ACDC [<numeric value>] [ ,@1]
:DCAC [<numeric value>] [ ,@1]
Description : Set the primary and secondary display to current measurement function for
the DCA, ACA or AC+DC mode. The <numeric value> parameter is used to
configure the measurement range. [ [@1] | ,@2 ] has the same syntax as
SCPI <channel_list> syntax. Use @1 to select the primary display function,
and @2 to select the secondary display function.
Parameter : <numeric value> ; For current range : 0.012, 0.04, 0.12, 1.2, 12 A.
[
[ ,@1] | ,@2 ] ; For the primary and secondary display setting.
Example : CONF:CURR:DC ;Set the primary display to DC current.
CONF:CURR:AC 0.12,@2 ;Set the
secondary display to AC current 120mA
range.
CONF:CURR:ACDC ,@1 ;Set the primary display to AC+DC
current.
CONFigure[:SCALar]:DIOCtest