CONFigure[:SCALar]:OFFDual
Description : Turn off the secondary display.
CONFigure[:SCALar]:OFFRecord
Description : Exit the dynamic recording mode ( MAX / MIN ).
CONFigure[:SCALar]:RANGe:AUTO <Boolean> [ [ ,@1] | ,@2 ]
Description : Enable or disable the auto range mode.
Parameter : <Boolean> ( 0n,1; off,0 )
CONFigure[:SCALar]:RANGe:AUTO? [ [ ,@1] | ,@2 ]
Description : This queries return the auto range status.
Response : 1 ( enable ) or 0 (disable ).
CONFigure[:SCALar]:RANGe:DIRection < UP | DOWN > [ [ ,@1] | ,@2 ]
Description : Change measurement range, increases it by one step.
Parameter : < UP >, increases it by one step ; < DOWN >, decreases it by one step.
CONFigure]
[:SCALar]
:RANGe? [
[ ,@1] | ,@2 ]
Description : Return the measurement range. Use @1 to select the primary display ,and
@2 to select the secondary display.
Response : Return the range value as Table 7-4.
Example : CONF:RANG? ; Query the primary display measurement range.
CONF:RANG? ,@2 ; Query the secondary display measurement