57
Configure List
Follow the steps below to configure list operation:
1. From the front panel, press and then (or press ). The following
display will show:
2. Select Edit and press . The following display will show:
3. Select either High-Rate or Low-Rate. These options configure the settable slew rate
range. If High-Rate is selected, users can adjust the slew rate of the transient in the
A/us range. If Low-Rate is selected, users can adjust the same in A/ms range. The
settable range will vary depending on the model.
4. The load will then prompt to enter a value for Current Range. Use the numeric keypad
or rotary knob to set the value for current range, then press to confirm.
5. The load will then prompt to enter a value for File Step, which is the total number of
steps for the list. Use the numeric keypad or rotary knob to set the value, then press
to confirm. Valid range is 2 – 84.
6. The load will prompt to enter values for each steps parameters, starting with the step
level. Use the numeric keypad or rotary knob to enter a value.