51
CALCulate subsystem
CALCulate:TOLerance:STATe < ON | OFF >
Description: Enable or disable tolerance mode
Example: CALCulate: TOLerance:STATe ON
CALCulate:TOLerance:STATe?
Description: Query the tolerance mode
Return: <ON, OFF >
CALCulate:TOLerance:NOMinal?
Description: Query the nominal value
Return: NR3 or -----(exceeding data range)
CALCulate:TOLerance:VALUe?
Description: Query the percentage value of tolerance
Return: NR3 or ----- (exceeding data range)
CALCulate:TOLerance:RANGe < 1 | 5 | 10 | 20 >
Description: Set tolerance range as 1%,5%,10% or 20% (20% not available for some models )
Example: CALCulate:TOLerance:RANGe 1
Set the tolerance range to 1%
CALCulate:TOLerance:RANGe?
Description: Query the tolerance range
Return: <BIN1, BIN2, BIN3, BIN4 or ---- >
---- means unset bin
CALCulate:RECording:STATe < ON | OFF >
Description: Enable or disable recording function
Example: CALCulate:RECording:STATe ON
CALCulate:RECording:STATe?
Description: Query the recording state
Return: <ON or OFF>
CALCulate:RECording:MAXimum?
Description: Query the maximum value of recording function
Return: <NR3, NR3> ( primary and secondary parameters, when data exceeds limits or there is
no data, what returns is----.)
CALCulate:RECording:MINimum?
Description: Query the minimum value of recording function
Return: <NR3, NR3> ( primary and secondary parameters, when data exceeds limits or there is
no data, what returns is----.)