EasyManua.ls Logo

BK Precision 9205B - List Commands

BK Precision 9205B
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
44
Command syntax: FETCh:DVM[:DC]?
Return parameter: <NRf>
MEASure:STATus?
This command is used to switch the display of VFD board between actual output state and DVM
state.
Command syntax: MEASure[:SCALar]:STATus <DVM|NORMAL>
5.9 LIST Commands
LIST:FUNCtion
This command is used to set the state of list mode.
Command syntax: [SOURce:]LIST:FUNCtion <0|1>
Query syntax: [SOURce:]LIST:FUNCtion?
Return parameter: 0|1
LIST:VOLTage
This command is used to set the steps of list file and voltage of current step.
Command syntax: [SOURce:]LIST:VOLTage <NRf>
Parameter: MIN to MAX
Unit: V
Query syntax: LIST:VOLTage?
Return parameter: <NRf>
Example: LIST:VOLT 1,3V //set the first step’s voltage to be 3 V
LIST:VOLT? 1 //query the first step’s voltage
LIST:CURRent
This command is used to set the steps of list file and current value of current step.
Command syntax: [SOURce:]LIST:CURRent <NRf>
Parameter: MIN to MAX
Unit: A
Query syntax: LIST:CURRent?
Return parameters: <NRf>
Example: LIST:CURR 1,2A //set the first step’s current to be 2 A
LIST:CURR? 1 //query the first step’s current
LIST:TIMEr
This command is used to set the steps of list file and delay time of current step.
This command is used to set delay time of each step.
www.calcert.com sales@calcert.com1.888.610.7664
0
5
10
15
20
25
30

Table of Contents

Related product manuals