H-80 Series User Manual V1.0
6 / 46
FiO
2
Display Accuracy
± (5% + 5% set value)
Temp
Range: 25°C ~ 45°C
Temp Display Accuracy
± 2°C
Flow
Range: 2 L/min ~ 80 L/min
Margin of Error: ± (2 L + 10% set value)
Flow Display Accuracy
± (2 L + 10% set value)
sPEEP
Range: 0 ~ 20 hPa
Margin of Error: ± (1 hPa + 20% set value)
Humidity Compensation
Range: -3~3,-3、-2、-1、0、1、2、3, the higher the range, the higher the humidification
output
Flow Step Interval
Range: 15 L/min ~ 80 L/min
Tube
Length: 1.8 m
Margin of Error: ± 10%
The connection should not fall off under the 45 N tension.
Specification of Patient Connector
φ 18 mm cone joint
6. Available Therapies
The device delivers the following therapies:
HFlow
– Under this mode, the setting range of flow is 15 L/min-80 L/min.
LFlow
– Under this mode, the setting range of flow is 2 L/min-25 L/min.
AutoFlow
– Under this mode, the flow cannot be set, output flow is greater than 15 L/min
and peak expiratory flow rate and the end-expiratory pressure is not lower than sPEEP
setting; applicable to H-80AS.
SmartFlow
– Under this mode, the setting range of flow is 15 L/min-80 L/min, the flow rate
when inhaling is the set value and the end-expiratory pressure is not lower than sPEEP
setting; applicable to H-80AS.