OZONE ANALYZER BMT 965 Manual, Rev. 01/2024
___________________________________________________________________________
41
M965_001_EN_2023
-
11
-
01
Cmd
Meaning
PC
-
>
965
965
-
>PC
9 Concentration?
float: Concentration
byte: Unit:
0: g/Nm
3
1: %wt/wt
2: ppm
v
3: g/m
3
4: ppm
w
5:
%wt
*9# *9#float,byte
10 Pressure?
= float * Multiplier
float: Pressure always in bar
byte Multiplier:
0: 1.0 (Unit bar)
1: 14.50778 (Unit psi)
2: 750.0617 (Unit Torr)
3: 0.1
(Unit MPa)
*10# *10#float,byte
11 Temperature?
float: in Kelvin
*11# *11#float
12
Operating hours?
*12#
*12
#word
13 High Alarm Parameter?
(first retrieve concentration unit)
float: Limit in present unit
byte1: Enabled
byte2: Latching
*13# *13#float,byte1, byte2
14 Low Alarm Parameter?
(first retrieve concentration unit)
float: Limit in present unit
byte1: Enabled
byte2: Latching
*14# *14#float,byte1,
byte2
15 Set High Alarm Limit (High Limit > Low Limit!)
(first retrieve concentration unit)
*15#float *15#
16 Set Low Alarm Limit (High Limit > Low Limit!)
(first retrieve concentration unit)
*16#float *16#
17 Set High Alarm Latching
byte 1: Latching 0: not Latching
*17#byte *17#
18 Set Low Alarm Latching
byte 1: Latching 0: not Latching
*18#byte *18#
19 Set High Alarm Enabled
byte 1:
Enabled 0: not Enabled
*19#byte *19#
20 Set Low Alarm Enabled
byte 1: Enabled 0: not Enabled
*20#byte *20#
21 Normalizing temperature?
float: in Kelvin
*21# *21#float