OZONE ANALYZER BMT 965 Manual, Rev. 01/2024
___________________________________________________________________________
42
M965_001_EN_2023
-
11
-
01
Cmd
Meaning
PC
-
>
965
965
-
>PC
23 Normalizing pressure?
= float x Multiplier
float: Pressure always in bar
byte Multiplier:
0: 1.0 (Unit bar)
1: 14.50778 (Unit psi)
2: 750.0617 (Unit Torr)
3: 0.1
(Unit MPa)
*23# *23#float,byte
29 Time?
hh,mm,ss
*29# *29#byte,byte,
byte
30
Set Hour
*30#byte
*30#
31
Set Minute
*31#byte
*31#
32
Set Second
*32#byte
*32#
33 Date Format?
0: DD.MM.YY
1: MM/DD/YY
*33# *33#byte
34 Set Date Format
0: DD.MM.YY
1: MM/DD/YY
*34#byte *34#
35 Date?
DD,MM,YY
*35# *35#byte,byte,byte
36
Set Day
*36#byte
*36#
37
Set Month
*37#byte
*37#
38
Set Year (YY)
*38#byte
*38#
39 RS-232 Timed/Polled?
1: Timed
0: Polled
*39# *39#byte
40 Set RS-232 Timed/Polled
1: Timed
0: Polled
*40#byte *40#
41 RS-232-Interval? (if Timed)
byte:
Seconds
*41# *41#byte
42 Set RS-232-Interval (if Timed)
byte:
Seconds
*42#byte *42#
43 Set Analog Output (Simulation)
Range 0.0: 0 V/ 4mA... 1.0: 10V /20 mA
2.0: normal operation
*43#float *43#
44 Autozero Interval?
byte: hours
if
0: no Autozero
*44# *44#byte
45 Set Autozero Interval
byte: hours
if 0: no Autozero
*45#byte *45#