OZONE ANALYZER BMT 965 Manual, Rev. 01/2024
___________________________________________________________________________
43
M965_001_EN_2023
-
11
-
01
Cmd
Meaning
PC
-
>
965
965
-
>PC
46 Alarm Beep?
1: On
0: Off
*46# *46#byte
47 Set Alarm Beep
1: On
0: Off
*47#byte *47#
48 Cuvette Status?
float: % Dirty
0: Clean
*48# *48#float
83 Trigger Zero (Parameter for security)
float: Dirty
If Autozero Interval > 0
, this
command will take about 20 s
*83#3.14159 *83#float
85
Firmware
-
Version?
*85#
*85#float
86 Status (decimally coded, as described in table on page 28,
BMT 964
-
legacy command
)
*86# *86#short
91 Set Link-Mode Timeout (after switching on always 10 s)
byte: Seconds
*91#byte *91#
93 Alarm Relays closing on Alarm?
1: closing
0: opening
*93# *93#byte
94 Set Alarm Relay Mode
1: closing
0: opening
*94#byte *94#
95 Set User Baud Rate
0: 2400
1: 4800
2: 9600
3: 19200
4: 38400
*95#byte *95#
98
Reset all parameters to standard factory setting
*98#
*98#
99 Set PIN (prevents changing parameters)
0: no PIN protection
*99#short
(0..9999)
*99#
101 Set Purge Time
byte: Seconds (10..100s)
*101#byte *101#
102 Get Purge Time
byte: Seconds
*102# *102#byte
114 Get Concentration Log Interval
short: seconds
*114# *114#short
115
Set
Concentration Log Interval
*115#short
*115#
116 Concentration Log Size (sum of both logs)?
word: byte
*116# *116#word
117 Error Log Size?
word: byte
*117# *117#word
118 Event Log Size?
word: byte
*118# *118#word
119
Delete Binary
Concentration Log
*119#
*119#
120
Delete CSV
-
Concentration Log
*120#
*120#