OZONE ANALYZER BMT 965 Manual, Rev. 01/2024
___________________________________________________________________________
43
46 Alarm Beep?
1: On
*46# *46#byte
47 Set Alarm Beep
1: On
*47#byte *47#
48 Cuvette Status?
float: % Dirty
*48# *48#float
83 Trigger Zero (Parameter for security)
float: Dirty
command will take about 20 s
*83#3.14159 *83#float
86 Status (decimally coded, as described in table on page 28,
*86# *86#short
91 Set Link-Mode Timeout (after switching on always 10 s)
*91#byte *91#
93 Alarm Relays closing on Alarm?
1: closing
*93# *93#byte
94 Set Alarm Relay Mode
1: closing
*94#byte *94#
95 Set User Baud Rate
0: 2400
1: 4800
2: 9600
3: 19200
*95#byte *95#
Reset all parameters to standard factory setting
99 Set PIN (prevents changing parameters)
*99#short
(0..9999)
*99#
101 Set Purge Time
*101#byte *101#
102 Get Purge Time
*102# *102#byte
114 Get Concentration Log Interval
*114# *114#short
Concentration Log Interval
116 Concentration Log Size (sum of both logs)?
*116# *116#word
117 Error Log Size?
*117# *117#word
118 Event Log Size?
*118# *118#word