OZONE ANALYZER BMT 965 Manual, Rev. 01/2024
___________________________________________________________________________
44
121 Get Errors (full, binary coded):
Bit 0: Lamp Low Warning
Bit 1: Lamp Low Error
Bit 2: Lamp Off Error
Bit 3: Cuvette Dirty Warning
Bit 4: Cuvette Dirty Error
Bit 5: Overpressure Error
Bit 6: Overrange Error
Bit 7: EEPROM Error
Bit 8: Lamp High Error
Bit 9: not used
Bit 10: not used
Bit 11: SD Card Warning
Bit 12: Low Concentration Warning
Bit 13: High Concentration Warning
Bit 14: Low Pressure Error