5 ELECTRICAL CONNECTIONS
5.1 BASIC INFORMATION
- It is recommended that the power supply negative pole be grounded.
- It is possible to supply up to eight 350 ohm load cells or sixteen 700 ohm load cells.
- Connect terminal “0 VDC” to the RS485 common of the connected instruments in the event
that these receive alternating current input or that they have an optoisolated RS485.
- In case of an RS485 network with several devices it is recommended to activate the 120
ohm termination resistance on the two devices located at the ends of the network, as
described in the paragraph RS485 SERIAL CONNECTION.
5.2 WIRING DIAGRAM
2 outputs: settable setpoints or remote output management via protocol.
2 inputs (Default: SEMI-AUTOMATIC ZERO input 1; NET/GROSS input 2): settable to
have the following functions: SEMI-AUTOMATIC ZERO, NET/GROSS, PEAK or REMOTE
CONTROL (see paragraph OUTPUTS AND INPUTS CONFIGURATION).