11 ALARMS

: the load cell is not connected or is incorrectly connected; the load cell signal
exceeds 39 mV; the conversion electronics (AD converter) is malfunctioning;
 
: the weight display exceeds 110% of the full scale.
 
: internal instrument converter failure; check load cell connections, if necessary
contact Technical Assistance.

: the weight exceeds the maximum weight by 9 divisions.
 
: maximum displayable value exceeded (value higher than 999999 or lower than -
999999).

: weight too high: zero setting not possible.

: this message appears in the sample weight setting, in real calibration, after the
eighth sample weight value has been entered.

: the value set for the parameter is beyond the permitted values; press to quit
the setting mode leaving the previous value unchanged. Examples: a number of
decimals is selected for full scale which exceeds the instrument's display potential;
value above the maximum setting value; the weight value set in sample weight
verification does not match the detected mV increase; the analog output correction
goes beyond the permitted limits.

: lock active on menu item, keypad or display.

: It’s not possible to display properly the number because is greater than 999999 or
less than -999999.

: gross weight equal to zero: the semi-automatic tare operation cannot be
performed.
Serial protocols alarms:

 
 

 

MODE
Bit LSB
76543210
xxxxxxx1
76543210
xxxx1xxx
76543210
xxxxxx1x
76543210
xxxxx1xx
76543210
On gross:
xxx1xxxx
On net:
xx1xxxxx
The response
to the zero
command is a
'value not
valid' error
(error code 3)
Status
Register
MODBUS
RTU
ASCII
__O-F_
__O-L_
__O-F_
__O-L_
__O-F_
&aa#CR
RIP *
__O-F_
__O-L_
__O-F_
__O-L_
__O-F_
__O-F_
HDRIP-N
_ERCEL
_ER_OL
_ER_AD
######
_ER_OF
O__SET
CONTIN
_ERCEL
_ER_OL
_ER_AD
^^^^^^
_ER_OF
O__SET
* For RIP remote displays, if the message exceeds 5 digits the display reads

.
If an alarm becomes active the relays open and the analog outputs go to the lowest
possible value according to the following table:
RANGE
0/20 mA
4/20 mA
0/5 V
0/10 V
-10/10 V
-5/5 V
Output value
-0.2 mA
3.5 mA
-0.5 V
-0.5 V
0 V
0 V