User manual WTS Ver.1.16
Seite 33 von 52
11 ALARMS
: the load cell is not connected or is incorrectly connected; the load cell signal
exceeds 39 mV; the conversion electronics (AD converter) is malfunctioning;
: the weight display exceeds 110% of the full scale.
: internal instrument converter failure; check load cell connections, if necessary
contact Technical Assistance.
: the weight exceeds the maximum weight by 9 divisions.
: maximum displayable value exceeded (value higher than 999999 or lower than -
999999).
: weight too high: zero setting not possible.
: this message appears in the sample weight setting, in real calibration, after the
eighth sample weight value has been entered.
: the value set for the parameter is beyond the permitted values; press to quit
the setting mode leaving the previous value unchanged. Examples: a number of
decimals is selected for full scale which exceeds the instrument's display potential;
value above the maximum setting value; the weight value set in sample weight
verification does not match the detected mV increase; the analog output correction
goes beyond the permitted limits.
: lock active on menu item, keypad or display.
: It’s not possible to display properly the number because is greater than 999999 or
less than -999999.
: gross weight equal to zero: the semi-automatic tare operation cannot be
performed.
Serial protocols alarms:
xxxxxxx1
xxxx1xxx
xxxxxx1x
xxxxx1xx
On gross:
xxx1xxxx
On net:
xx1xxxxx
to the zero
command is a
'value not
valid' error
(error code 3)
Register
MODBUS
RTU
* For RIP remote displays, if the message exceeds 5 digits the display reads
.
If an alarm becomes active the relays open and the analog outputs go to the lowest
possible value according to the following table: