User manual WTS Ver.1.16
Seite 29 von 52
-

: continuous weight transmission protocol to RIP6100, RIP675, RIP6125C
series remote displays; the remote display shows the net weight or gross weight
according to its settings (set:

=

,

=

,

=
).
-

: continuous weight transmission protocol to RIP6100, RIP675, RIP6125C
series remote displays (set:

=

,

=

,

=
).
When the remote display is set to gross weight:
- if the instrument displays the gross weight, the remote display shows the gross
weight.
- if the instrument shows the net weight the remote display shows the net weight
alternated with the message ‘

'.
-

: continuous weight transmission protocol to RIP675Y series remote displays
(set:

=

,

=

,

=
, the settings cannot be changed).
-

: transmission speed (2400, 4800, 9600, 19200, 38400, 115200; default:
9600).
-

: instrument’s address (from 1 to 99; default: 1).
-

: maximum transmission frequency (10 20 30 40 50 60 70 80;
default: 10); to be set when the

transmission protocol is selected.
Maximum setting frequency (

):
- 20Hz with minimum baud rate 2400 baud.
- 40Hz with minimum baud rate 4800 baud.
- 80Hz with minimum baud rate 9600 baud.
-

: delay in milliseconds which elapses before the instrument replies
(from 0 to 200 msec; default: 0).
-

:
-

: parity none (default).
-

: even parity.
-

: odd parity.
-

: stop bit (1 2; default: 1).