Precision™ Spinal Cord Stimulator System Clinician Manual
Clinician Manual
97035873-01 70 of 75
NOTE These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reection from structures, objects and people.
a Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to xed RF transmitters, an electromagnetic site survey
should be considered. If the measured eld strength in the location in which the Spectra WaveWriter Spinal Cord
Stimulator System is used exceeds the applicable RF compliance level above, the Spectra WaveWriter Spinal Cord
Stimulator System should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the Spectra WaveWriter Spinal Cord Stimulator
System.