Electromagnetic Compatibility
Clinician Manual
97035873-01 71 of 75
Recommended separation distances between portable and mobile RF communications
equipment and the Precision System
The Precision System is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Precision System can help prevent
electromagnetic interference by maintaining a minimum distance of 30 cm between portable and mobile
RF communications equipment (transmitters) and the Precision System.
NOTE These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reection from structures, objects and people.