3 Functionality
900-350_IFU_V2.6_10296-S13-20100111-EN ARC 300 / ARC 350 Operating manual 29
3.3. Activation and alarm signals in monopolar and
bipolar mode
The activation and alarm signals are emitted at volume level 19. The
mnimum volume is level 16. If required, the volume of the activation
signals can be adjusted for a louder environment. The alarm sound
cannot be changed.
Mode Indicator in
menu program
Frequency (Hz) Signal type
Monopolar Cut SOUND 1 700 Continuous sound
Monopolar Coag SOUND 2 500 Continuous sound
Bipolar Cut SOUND 3 550 Continuous sound
Bipolar Coag SOUND 4 340 Continuous sound
Sim Coag
Forced Cut
SOUND 5 800 Continuous sound
Ligation End Pulsating
alternating sound
Alarm – 1000 Intermittent
Pos: 18 /679-BOWA/ARC 300/35 0/4 Funktionsweise/ 4.4 Not-Aus @ 4\mod_1261 040463116_6.doc @ 34691 @ 2
3.4. Emergency shutoff
The HF device can always be switched off in an emergency using the
on/off switch 33.
Pos: 19 /679-BOWA/ARC 300/35 0/4 Funktionsweise/ 4.5 Überwachungsfunk tionen @ 4\mod_12610404 72366_6.doc @ 34710 @ 233
3.5. Monitoring functions
3.5.1. Self-test
When the HF device is switched on, it runs a self-test that checks the
operating elements, acoustic signal, microprocessor and hardware
function.
3.5.2. Cyclical test during operation
During operation, safety-relevant functions and signals will be tested
cyclically. If errors are detected, the HF generator will shut itself off. An
error message appears on the display. For further information, see
Detecting and correcting errors, Page 64.