8. Preparation
Pos: 43 /679-BOWA/ARC 300/35 0/9 Aufbereitung/ 9.1 Aufbereitung des Zubehörs/Desinfizier en und Reinigung @ 4\mod_12 61040694772_6.doc @ 35185 @ 22
8.1. Preparation of the accessories
X Prepare the accessories (e.g. surgical handles, instruments, active
electrodes, neutral electrodes and cables) as described in the
corresponding operating manuals.
X Check the accessories before and after use for damage and to
ensure that they are working properly.
8.2. Disinfection and cleaning
NOTE
Incorrect handling of the HF device can cause damage to the unit!
X Never sterilize the HF devices ARC 300/350. Clean or disinfect
them instead.
WARNING
Risk of electric shock and fire!
X Unplug the power connection before cleaning the device.
X For cleaning surfaces, use the approved cleaning
agents/disinfectants only as specified by the manufacturer.
X Ensure that no liquid penetrates the device.
X Ensure that the AUTOSTART function is deactivated.
1. Apply the cleaning agent and disinfectant.
2. Wipe the agent off with a sponge moistened with clean water or
with a cloth.
3. Dry the device using a clean, lint-free cloth.