7 Detecting and correcting errors
68 ARC 300 / ARC 350 Operating manual 900-350_IFU_V2.6_10296-S13-20100111-EN
7.2. Errors in the EASY monitoring
When working with a split neutral electrode, the following errors may
occur:
EASY monitoring
20
Cause Indicator Corrective measures
Flashes green Significant increase in
resistance
–
Application must not be
terminated.
When the monopolar current
is activated, a significant
problem occurs
An acoustic signal sounds.
A corresponding “INF” - "xxx"
message appears on the 7-
segment indicator
“Monopolar Cut” 2 and the 7-
segment indicator
“Monopolar Coag” 9.
X Check the neutral
electrode and the
neutral electrode
cable; see Chapter
EASY neutral
electrode monitoring
(EASY monitoring),
Page 30.
X Check the neutral
electrode cable for
proper contact and
external damage.
Switches from green to
continuous red
Loosening electrode An acoustic signal sounds.
The corresponding “INF” –
“xxx” message appears.
The device's electronics
system switches off the
monopolar outputs of the
connection sockets 36 and
40.
X Correct the position of
the neutral electrode.
In the case of
continuing error
messages, replace
the neutral electrode.