7 Detecting and correcting errors
900-350_IFU_V2.6_10296-S13-20100111-EN ARC 300 / ARC 350 Operating manual 67
Error no.
(INF)
Signa-
ls/System
events
Cause Corrective measures
149
3-pin plug of the
handpiece has been
inserted incorrectly
Handpiece possibly
defective
Monopolar 4 mm
connection cable has
been connected to output
2 40 in the right-hand 4
mm connection socket
X
Check the handpiece.
X Replace the handpiece.
X Connect the 4 mm connection cable
to output 2 40 in the left-hand 4 mm
connection socket.
X Contact the service department if an error cannot be corrected with
the measures listed; see Chapter Technical service, Page 72.
Errors not listed
X Contact the service department regarding errors that are not shown
in the error list; see Chapter Technical service, Page 72.
X If the expected change in the tissue does not occur and no error
message appears, check the parameters and the accessory
connections.