6 Operation
46 ARC 300 / ARC 350 Operating manual 900-350_IFU_V2.6_10296-S13-20100111-EN
DANGER
Blood loss due to incomplete blood vessel sealing!
X Deactivate the ligation instrument after completed sealing of blood
vessels.
;
The key/display for "Monopolar output 1" 35 and "Bipolar output 3"
44 are preconfigured.
X Connect the ligation instrument to "Bipolar output 3" 43 of the HF
device.
6.3.5. Program 4 “Argon Ligation”
(ligation only for ARC 350L)
If the software option “Ligation” is unavailable, this program provides the
function “Standard Bipolar”.
In this program, open surgical interventions are carried out with the
additional device ARC PLUS for argon electrocoagulation. When
suitable instruments are connected, argon coagulation can be carried
out with rigid electrodes and argon cutting.
This program is used in a wide range of surgical applications, e.g.
gynecology and laparoscopy.
It facilitates permanent blood vessel sealing with the manual
instruments of the BOWA NightKNIFE
®
, LIGATOR
®
, and TissueSeal
®
series for large veins, arteries or tissue bundles.
Intended for:
• large-lumen veins, arteries or tissue bundles
• use in combination with instruments of the
BOWA NightKNIFE
®
, LIGATOR
®
, and TissueSeal
®
series
Not intended for:
• tubal sterilization or sterilization by means of tubal
coagulation
• dirty or defective electrode tips