You can change the following factory settings:
Select Cutting settings
1. Press the key/display for „Monopolar Output 1“ 35 or „Monopolar
Output 2“ 39 and select connection socket 1 or 2.
2. Press the power limitation key for monopolar cutting 3 or 4.
–Or–
1. Press the key/indicator for "Bipolar Output 3" 44 or "Bipolar
Output 4" 48.
2. Press the power limitation key for bipolar cutting 23 or 24.
Set blend factor
1. Press the key for the monopolar degree of scabbing 5/6.
–or–
Press the key for the bipolar degree of scabbing 25/26.
2. Set the blend factor between "1-9" or to "0" for coagulation-free
cutting.
Select current type for coagulation current
1. Press the key/indicator for "Moderate Coagulation" current 12
– or –
Press the key/indicator for "Forced Coagulation" current 13
– or –
Press the key/indicator for "Spray Coagulation" current 14.
2. Set the power limit for the monopolar coagulation current using the
keys for power limitation for monopolar coagulation 10/11.
–Or–
1. Set the power limit for the bipolar coagulation current using the
keys for power limitation for bipolar coagulation 30/31.