11 Technical specifications
900-350_IFU_V2.6_10296-S13-20100111-EN ARC 300 / ARC 350 Operating manual 75
Dimensions and weight
External dimensions: width x height x depth (mm) 430 x 180 x 400
Weight Approx. 14 kg
Programs
Number of programs in the device 100
Standard programs, factory set √
Individually programmable √
Program number and information shown on the
display
√
Monitoring of the neutral electrode
EASY: Electrode Application System √
Indicator for solid or split electrode on the front
panel
√
Indication of the transition resistance between the
section surfaces of the split
√
neutral electrode on the display √
Indication of the line resistance when using a solid
neutral electrode on the display
√
Max. resistance between the section surfaces of
split electrodes
250 Ω
Warning signal in the case of risk under the
neutral electrode
Visual, acoustic
Warning indicator as a text on the display: √
Safety features
ISSys: Integrated safety system √
Continuous monitoring of the HF leakage current
with error messages
√
Monitoring of the dosage, error message on the
display
√
Continuous self-test √
Continuous status indication on the display
√
Display of operating errors on the display √
Display of system errors on the display √