11 Technical specifications
900-350_IFU_V2.6_10296-S13-20100111-EN ARC 300 / ARC 350 Operating manual 93
Setting Program
Forced COAG Non-cutting, Cutting, Mixed 0-6, 9, 12, 16, 20,
11 (31 W and above),
23 (only socket 1),
24-99 (socket 1+2)
ARC350FCOAG_SE_080124_1
0
20
40
60
80
100
120
140
10 100 1000 10000
P [W]
Forced Coag
P=60W
Forced Coag
P=120W
Measurement of resistances
• Diagram of power output P [W] as a
function of the load resistance R [Ω] for
the setting “Forced COAG Mixed, Non-
cutting, Cutting” = 120 W/60 W
ARC350FCOAG_SE_080124_2
0
20
40
60
80
100
120
0 20 40 60 80 100 120
displayed value
P [W]
• Diagram of power output P [W] as a
function of the "Forced COAG Mixed,
Non-cutting, Cutting" setting. Measured
load resistance = 500 Ω
0
500
1000
1500
2000
2500
3000
3500
4000
4500
5000
02
10
0108
060
40
2
0
displayed value
U [Vp]
ARC350FCNC_SE_080125
• Diagram of HF output voltage U [Vp] as
a function of the “Non-cutting Forced
COAG” setting (idle mode)
0
500
1000
1500
2000
2500
0210
0
1
08
06
0
4
0
2
0
displayed value
U [Vp]
ARC350FCC_SE_080125
• Diagram of HF output voltage U [Vp] as
a function of the “Cutting Forced
COAG” setting (idle mode)
ARC350FCOAG_SE_080313_3
0
500
1500
1000
2000
2500
3000
3500
0 20 40 60 80 100 120
displayed value
U [Vp]
• Diagram of HF output voltage U [Vp] as
a function of the “Mixed Forced COAG”
setting (idle mode)