3.5 Troubleshooting: Camera Cart
Camera Occurrences
There are several possible occurrences that can be attributed to communication failures related to
the IR camera.
Occurrence Symptoms Possible Causes
Decrease in cam-
era volume
Restricted field of view
for navigation
LEDs in camera lens
are not on
Overheating
Communication error between the Monitor
and Camera Carts
IR light from another device causing interfer-
ence (e.g., pulse oximeter).
Surfaces in the OR are unintentionally reflect-
ing IR light (e.g., reflective strips on clothing)
No power
Camera does not beep
during boot-up
Problem with internal cabling
Problem with external cabling
No communication
Camera does not beep
during application start up
Communication error between Monitor and
Camera Carts: Check cabling
Not tracking mark-
er spheres proper-
ly
Software displays uniden-
tifiable gray marker
spheres in the camera
field of view
IR light from another device causing interfer-
ence (e.g., pulse oximeter).
Surfaces in the OR are unintentionally reflect-
ing IR light (e.g., reflective strips on clothing)
Camera motor
control does not
work
Pressing the motorized
camera control buttons
does not move camera
Motor on/off switch is turned off: Turn switch
ON
Communication error: Check cabling
If green camera status LED is off, communi-
cation is not possible: Reboot device
NOTE: If you are having problems with your camera, contact Brainlab support.
Troubleshooting: Camera Cart
42 Technical User Guide Rev. 1.1 Curve Dual Navigation Station Ver. 1.2