5.5 Power Plugs
General Information
The Curve system comes equipped with a specific power plug suitable for the region of use.
Use of Specified Power Cords
Only use power cords as specified by Brainlab. Do not extend power cords.
Reliable grounding can only be achieved when the system is connected to an equivalent
receptacle marked “Hospital Only” (“Hospital Grade” in North America). Only insert the
main plug into a socket outlet that is protectively earthed. Do not negate the protective
action by the use of an extension cable.
Power Plugs
78 Technical User Guide Rev. 1.1 Curve Dual Navigation Station Ver. 1.2