Test Step Instructions and Conditions
Check and prepare
for normal use
After testing, check that the Curve system is restored to the conditions
necessary for normal use before being returned into service.
Remove all devices that have been connected, e.g., measurement lines.
Critical Values
Test Step Normal Condition Single Fault Condition
Touch Current
< 100
µA
500
µA
Reference Dimension
Use the table starting on page 60 to enter the single measured values at each required
measurement point.
Enter the maximum measured value in the table below:
Test Step Normal Condition Single Fault Condition Passed?
Touch Current
Serial number of measuring device: ________________________
Calibration valid through (date): ________________________
Test performed (date):_________________by: __________________________________
ELECTRICAL SAFETY
Technical User Guide Rev. 1.1 Curve Dual Navigation Station Ver. 1.2 59