52 67XXU USB/Serial Scale Service Manual
LINE CAL.P0 Line calibration point0 Do zero point calibration. This point
can't be omitted.
CAL.P1 Line calibration point1 First weight point calibration. This
point can't be omitted and standard
weight must be over 10%FS.
END.Y YES End calibration? YES=go to CAL.END to end
NO=go to do next point calibration
NO
CAL.P2 Line calibration point2 Second weight point calibration.
Standard weight must be over
10%FS and be larger than it in
CAL.P1. This point can be omitted.
END.Y YES End calibration? YES=go to CAL.END to end
NO=go to do next point calibration
NO
CAL.P3 Line calibration point3: Third weight point calibration.
Standard weight must be over
10%FS and be larger than it in
CAL.P2, this point can be omitted.
geo code 00-70 Select geograhical position Code 00-70
gravt 9.76183-
9.99999
Input gravity of user location
input Input or view calibration
parameter values
Only used to copy calibration
parameters from one scale to a new
scale.
CAL