72 67XXU USB/Serial Scale Service Manual
7.6.4 Parameter Multi
Set the USER-COM-LYOUT parameter to mutli.
Output string frame
Command: W<CR> (57h 0dh), request current reading
<LF><Prompt><p>W
1
W
2
W
3
W
4
W
5
<DP>W
6
U
1
U
2
U
3
U
4
U
5
<CR>
Line number and content are determined by setting of USER-OUT-xxxx
<LF><Prompt>H
1
H
2
H
3
H
4
<CR>
USER-OUT-STATUS is set to YES
C<CR>
USER-OUT1/2-LINE is set to LINE1/2/3/4
-----
The number of lank lines is determined by USER-OUT-LINE setting
<LF>
(1) The decimal point position is determined by CONFG-PRIM.D
(2) The unit position and bytes are determined by which current unit is used
(3) The details of <Prompt> refer to the content in User submenu
(4) In HOLD mode, if ADC conversion speed is set to high speed (80Hz) and USER-COM-
LYOUT is set to MULTI, the output from COM may be slower than the data processed by the
indicator. For "real time" data, select fewer output contents and set a higher baud rate for
C<CR> --- USER-OUT-LINE is set to LINE1/2/3/4
<LF> ---
---The number of blank lines is determined by USER-OUT1/2-LINE setting
<ETX>
--- Last byte of string frame