EasyManua.ls Logo

Brecknell 6702U - Page 77

Brecknell 6702U
86 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
67XXU USB/Serial Scale Service Manual 77
Command: <US> F (1Fh 46h)
Command: all others
Table 7.9 Summary of Command and Response:
Response
Scale is taken out of echo mode. All subsequent characters are commands.
<STX>F<CR>
Response
Unrecognized command
<STX>?[status]<CR>
Command Response
ASCII HEX
<US> W 1F 57 1. normal data
<STX>W
1
W
2
<dp>W
3
W
4
W
5
<CR>
2. if current weight is invalid
<STX>?<Status Byte><CR>
<US> Z 1F 5a <STX>?[status]<CR>
<US> H 1F 48 1. normal data
<STX>W
1
W
2
<dp>W
3
W
4
W
5
W
6
<CR> -- weight data (x10)
2. if current weight is invalid
<STX>?<Status Byte><CR>
<US> A 1F 41 <STX>?<CR>
<US> B 1F 42 <STX>[confidence]<CR?>
<US> E 1F 45 <STX>E<CR>
<US> F 1F 46 <STX>F<CR>
others <STX>?[status]<CR>

Table of Contents