4 Description of function
18 R-220 Pro Continuous Operation Manual, Version A
Heating bath with
heating liquid
Evaporating solvent
Rotation drive for
evaporating ask
Cooler with
cooling medium
Optimal condensation
zone range
(~ 80% cooler height)
Condensated solvent
a
b
c
Safety reserve
a
Evaporation area
The solvent is heated by means of a heating bath. The turbulent mixing inside the rotating
evaporating ask results in an increased evaporation rate.
The drive unit oers a wide range of rotational speeds to perfectly match dierent evaporation
tasks. It also leads to an even mixing of the sample, thus preventing stationary overheating in
the ask. When drying powders or pasty materials, the rotation ensures smooth and
thorough drying.
b
Cooling area
The solvent vapor quickly enters the condenser. Here, the energy in the solvent vapor is
transferred to the cooling medium (mostly water), so that the solvent condenses.
c
Receiving ask
The receiving ask collects the condensed solvent.
NOTE
For information about optimum distillation conditions see section 6, Operation.