5 Putting into operation
46 R-220 Pro Continuous Operation Manual, Version A
5.12 Heating bath setup for 2×1800 W heating element
The heating element holder is labelled with the heating power data. The heating bath of the 2×1800 W
version can be lled with dierent kinds of liquid heating transfer medium. Depending on the maximum
heating bath temperature for the target application, a suitable medium must be used.
How to set the maximum setpoint temperature
Limiting the maximal temperature set-point to a value below 180 °C can help to eliminate processing
errors, e.g. when working with heat-sensitive substances.
AERATE
Temp
+
Set Temp
SET
Store setting
X Keep the "SET" button pressed while switching on
the Rotavapor® to enter the setup menu.
X Set the temperature limit via the turning knob. The
set value is displayed as bath temperature value
on the display.
X Press "AERATE" button to store the setting and to
leave the menu.