5 Putting into operation
52 R-220 Pro Continuous Operation Manual, Version A
5.15 Water
With the heating bath in lowest position, install the level drain output tube with a level dierence of
> 5 cm towards the sink to ensure safe draining. Fixate all tube connections with hose clamps!
NOTICE
Risk of damage caused by heating transfer mediums.
Do not ll the heating bath to overow
Do not leave the system unattended when replenishment is active
Check the level drain for proper drainage function in lowest heating bath position
Do NOT connect the replenishment tap to a water source in case
a heating transfer medium dierent to water (e.g. glycerin) is used!
!
!
Sink (dummy)
Sink (dummy)
a
b
c
To condenser
a
Level drain with Ø 12 mm hose connector, let into sink.