5 Putting into operation
33 R-220 Pro Continuous Operation Manual, Version A
5.3 Sensor setup and adjustments
Sensor setup menu
These instructions are valid for the setup menu of the cooling water sensor and level sensors
connected at the rear connection eld.
AERATE
=
LEVEL1
AERATE
=
LEVEL2
AERATE
=
Lift
+
Flow
a
b
c
Store selection
Store selection
Store selection and
leave setup menu
X Keep the "LIFT UP" button pressed while
switching on the Rotavapor® to enter the setup
menu.
X Press the "UP" button to toggle between 'on'
(cooling water ow sensor activated) and 'o'
(sensor deactivated).
X Press "AERATE" button to store the selected
sensor setting and automatically proceed with
level sensor 1.
X Press the "UP" button to toggle between 'on'
(level sensor 1 activated) and 'o' (sensor deacti-
vated).
X Press "AERATE" button to store the selected
sensor setting and automatically proceed with
level sensor 2.
X Press the "UP" button to toggle between 'on'
(level sensor 2 activated) and 'o' (sensor deacti-
vated).
X Press "AERATE" button to store the selected
sensor setting and to leave the setup menu.